arxiv 1312.2744 v2 pith:BFNRBJDQ submitted 2013-12-10 physics.ins-det nucl-exphysics.med-ph

Calibration of photomultipliers gain used in the J-PET detector

classification physics.ins-det nucl-exphysics.med-ph
keywords photomultiplierscalibrationgainusedcurvedetectorelectricalj-pet
verification ladder T0 review T1 audit T2 compute T3 formal T4 reserved
0 comments
read the original abstract

Photomultipliers are commonly used in commercial PET scanner as devices which convert light produced in scintillator by gamma quanta from positron-electron annihilation into electrical signal. For proper analysis of obtained electrical signal, a photomultiplier gain curve must be known, since gain can be significantly different even between photomultipliers of the same model. In this article we describe single photoelectron method used for photomultipliers calibration applied for J-PET scanner, a novel PET detector being developed at the Jagiellonian University. Description of calibration method, an example of calibration curve and gain of few R4998 Hamamatsu photomultipliers are presented.